All Categories

Portable Gas Detector

FID - Flame Ionization Detector

FID - Flame Ionization Detector

This instrument enabling quantitative analysis of organic compounds based on the magnitude of the current signal. It features an extremely fast response speed, which makes it highly suitable for rapid screening and continuous monitoring across large field sites.
  • Overview
  • Videos
  • Resources
  • Related Products
Product Description
ItemDescription
Accuracy1.0~10,000 ppm: ±10% of reading or ±1.0 ppm (whichever is greater). This accuracy is based on a single-point calibration using methane at 100~500 ppm (including drift) at calibration temperature and humidity.3,000~50,000 ppm: ±10% of reading, using 20,000 ppm methane single-point calibration.
Repeatability±2% at 500ppm methane
Dynamic Range1.0~50,000 ppm methane
Linear Range1.0~50,000 ppm methane
Minimum Detection LimitDefined as 7 times the standard deviation of peak-to-peak noise: 0.5ppm methane
Response Time (T90)Using 10,000ppm methane, 20℃: ≤2.5 seconds
Recovery Time (T10)Using 10,000ppm methane, 20℃: ≤2.5 seconds
Sampling SpeedRated flow at probe inlet: 0.8L/min
BatteryType: Rechargeable Li-ion battery packCapacity: 5000mAhRated Voltage: 7.4VCharging Voltage: 8.4VFull Charge Time: 3 hours (power off state)Working Time (20℃): Approx. 9 hoursWarning: Do not replace the battery in a combustible gas environment or area with explosion threat!
Battery ChargerInput: AC 100~240VOutput: DC 8.4V ±0.05V, 2A ±0.2A (Constant Current)Indicator: Red (Charging), Green (Standby/Full)Protection: Over-current, over-voltage, short circuitWarning: Do not charge in a combustible gas environment or area with explosion threat!
FID Lifespan> 6000 hours under normal maintenance and cleaning
Hydrogen Capacity:10L (20℃)
Dimensions:30 × 108 mm
Weight:Approx. 100g
Working Temperature:-20℃~+60℃ (Hydrogen release decreases below 0℃)
Metal Hydride CylinderCharging Pressure: 0.7MPa @ 20℃Discharge Flow: 13ml/min (Max discharge flow: 300-500mL/min)Working Time (20℃): ≥10 hoursSafety: Low-pressure solid-state hydrogen storage, higher safety than high-pressure steel cylinders.
Housing MaterialHeat-resistant, chemical corrosion-resistant polymer material
Physical Dimensions (Host):322 mm (L) × 93 mm (W) × 61 mm (H)
Weight (Host):Single FID model: Approx. 0.90 kg; FID+PID dual detector model: Approx. 0.98kg
CommunicationSupports WIFI, Bluetooth
Data Storage16GB memory, capacity not less than 2 million detection data entries, adjustable storage interval.
InterfaceBuilt-in full Chinese/English interface, user switchable. Data export plugin also supports Chinese/English interface.
Hydrogen GeneratorPrinciple: Proton Exchange Membrane (PEM)Dimensions: 146 × 157 × 210 mmMass: 1.8kg ±5%Water Addition Type: Pure water or deionized waterWater Temperature: 20-65℃Charging Time: 2-4 hours (depending on temperature)